Our Technology, solution for Tomorrow too!

Measuring Instrument & Tester

Digital MΩ Tester

MODEL 3587

  • ■High speed sampling: 50 or 60 times/sec.
  • ■Test voltage: DC 25~1000 V (resolution: 1V)
  • ■Resistance range: 2/20/200/2000 MΩ
  • ■Connection check function available,prevents misjudgment became of bad connection
  • ■Compact & Light: 205(W)×64(H)×169(D)mm, 1kg

Catalog
Operation manual

Typical Application

Insulation test for general purpose electric instruments, transformers, batteries, etc.

Model Designation

3587 -
 
① Data Output
Code Description
X RS-232C
04 RS-232C + BCD (open collector)

Measuring Range

Test voltage Resistance range Accuracy
± (2% of rdg.+3 digit) ± 5% of rdg. ± 30% of rdg.
DC 25 V ~ DC 99 V 2.000 MΩ 0.000~2.000 MΩ - 2.010~4.990 MΩ
20.00 MΩ 1.80~20.00 MΩ - 20.10~49.90 MΩ
200.0 MΩ 18.0~100.0 MΩ 100.1~200.0 MΩ 201.0~999.0 MΩ
DC 100 V ~ DC 499 V 2.000 MΩ 0.000~2.000 MΩ - 2.010~4.990 MΩ
20.00 MΩ 1.80~20.00 MΩ - 20.10~49.90 MΩ
200.0 MΩ 18.0~100.0 MΩ 100.1~200.0 MΩ 201.0~499.0 MΩ
2000 MΩ - 180 ~2000 MΩ 2010 ~5000 MΩ
DC 500 V ~ DC 1000 V 20.00 MΩ 0.00~20.00 MΩ - 20.10~49.90 MΩ
200.0 MΩ 18.0~200.0 MΩ - 201.0~499.0 MΩ
2000 MΩ 180 ~1000 MΩ 1001 ~2000 MΩ 2010 ~9990 MΩ

Basic Specifications

Output voltage DC 25~1050 V (Resolution: 1 V),
Open voltage: ±(1% + 5 V) of output voltage
Open circuit voltage Less than 130 % of rated voltage
Rated current Less than 1 mA
Display Measured value: 0~9990
2.7 inch organic EL display (OLED), color: green
Sampling cycle
FAST:
50 times/sec. when the frequency setting is 50 Hz
60 times/sec. when the frequency setting is 60 Hz
SLOW:
10 times/sec.
Response speed
(During NG judgement)
Display response & judgement time:
Approximately 0.05s in case of fixed range.
Approximately 0.2s in case of automatic range.
Parameter hold Memorize and hold the parameter with EEPROM for 10 years.
Insulation resistance All terminals – External case: DC500V More than 100 MΩ
Withstand voltage All terminals – External Case: DC 2000 V 1min.
All terminals – External Case: DC 500V more than 100 MΩ
All terminals – Measurement terminals and control input/output terminals:
AC 1500 V 1min.
Measurement terminals – control input/output terminals and data input/output terminals: AC 1500 V 1min.
Power supply AC 100~240 V(50/60 Hz)
Power consumption 28 VA MAX.
Operating environment 0~50°C, less than 90% RH(no condensation)
Storage environment -20~70°C, less than 70% RH (no condensation)
Size 205(W)×64(H)×169(D)mm(not included protuberance)
Weight  
Accessories Power cord (1), Fuse (1), Operating manual (1)
Optional Accessories
5813-21
5813-21
LOW probe
5813-22
5813-22
HIGH probe
5813-23
5813-23
HIGH probe with switch +LOW probe
Power supply cord for AC 200V  5880-23-030
Panel mounting bracket  5811-31

page top