Digital Meter
DC Voltmeter / Ammeter / Receiver
MODEL 3194A
- ■ Maximum display 19999, Compact design of 48 x 24 mm
- ■ Scaling -19999 to 19999 is possible.
- ■ Offset fixing and scaling adjustment functions are equipped.
Standard Specifications
Model Designation
| 3194A | - | □ | - | □ | - | □ |
| ① | ② | ③ |
①Measuring Input
| Code | Measuring Range | Input Resistance | Accuracy * | Overload |
|---|---|---|---|---|
| 03 | ±1.9999 V | 1 MΩ | ±(0.05% of rdg. + 2 digit) | DC ±250 V |
| 04 | ±19.999 V | 1 MΩ | ±(0.05% of rdg. + 2 digit) | DC ±250 V |
| 05 | ±199.99 V | 10 MΩ | ±(0.05% of rdg. + 2 digit) | DC ±350 V |
| 09 | DC 1 to 5 V | 1 MΩ | ±(0.1% of rdg. + 5 digit) | DC ±250 V |
| V1 | DC 0 to 1 V | 1 MΩ | ±(0.05% of rdg. + 2 digit) | DC ±250 V |
| V2 | DC 0 to 5 V | 1 MΩ | ±(0.05% of rdg. + 2 digit) | DC ±250 V |
| V3 | DC 0 to 10 V | 1 MΩ | ±(0.05% of rdg. + 2 digit) | DC ±250 V |
| 19 | DC 4 to 20 mA | 13 Ω | ±(0.1% of rdg. + 5 digit) | DC ±150 m A |
| A1 | DC 0 to 1 mA | 200 Ω | ±(0.1% of rdg. + 2 digit) | DC ±40 mA |
- * Accuracy:
- Defined at 23℃ ± 5℃, 45 to 75% RH.
- Temperature coefficient:
- ±100 ppm/°C (-03, -04, -05, -V1, -V2, -V3)
±200 ppm/°C (-09, -19, -A1) Operating temperature range: 0 to 50°C
②Display Color
| Code | Content |
|---|---|
| R | Red LED |
| G | Green LED |
③Option
| Code | Function |
|---|---|
| X | none |
| A01 | External decimal point control |
General Specifications
| Display | 0 to 19999 red or green LED (character height 8mm) with zero suppression function |
|---|---|
| Scaling function | Full scale display: -19999 to +19999 (with full scale display setting function) Offset display: -19999 to +19999 (with offset display setting function) Adjustment: Display value can be adjusted. |
| Offset fixing function | Function of fixing display value to offset value when input is less than offset value. |
| 10⁰ digit fixed to 0 | Setting is possible. |
| Decimal point display | Setting is possible from the front panel, Terminal setting is also possibe for -A01(option). |
| Over-range display | Flashing with 130% display. (But when exceeded 19999, flashing is displayed by 0000.) |
| Hold function | Holds measured data (Not isolated from input). |
| Resolution | 1/20000 |
| Display cycle | Selectable from approx. 400 ms or approx. 1s |
| Input configuration | Single ended, floating input |
| A/D conversion | Δ-Σ conversion system |
| Noise rejection | Normal mode (NMR) – 50dB or more |
| Insulation resistance | DC 500 V 100 MΩ or more |
| Withstand voltage | Between input terminal and outer case: AC 1500 V 1minute Between power supply terminal and outer case: AC 1500 V 1minute Between power supply terminal and input terminal: AC 500 V 1minute |
| Power supply | DC 5 to 24 V |
| Power supply voltage tolerance | DC 4.75 to 32 V |
| Power consumption | DC 5 V approx. 150 mA DC 12V approx. 50 mA DC 24V approx. 35 mA |
| Operating ambient temperature | 0 to 50°C |
| Storage temperature | -20 to 65°C |
| Weight | Approx. 45 g |