Our Technology, solution for Tomorrow too!

Digital Meter

Thermocouple, Resistance Thermometer

MODEL 3196

  • ■ Compact size temperature meter
  • ■ Thermocouple input, RTD input available
  • ■ Adjustment function

Operation manual

Standard Specifications

Model Designation

3196 - - -

①Measuring Input

Thermocouple (T/C)
Code Temperature sensor Measuring Range Display Range Accuracy *1
01 R 100 to 1600°C -50 to 1750°C ±(0.5% of rdg. + 1°C)
02 K -199 to 1200°C -199 to 1350°C
03 E -199 to 900℃ -199 to 1050℃
04 J -40 to 750°C -199 to 1250°C
05 T -199 to 350°C -199 to 420°C
06 N -40 to 1200℃ -199 to 1350℃
07 B -600 to 1700°C -20 to 1810°C
*1 Accuracy:
Specified at measurement range
Defined at 23 ± 5°C, 45 to 75% RH.
Temperature coefficient:
±300 ppm/°C (Defined over an operating temperature range of 0 to 50°C)
Reference junction compensation:
± 2 °C for operating temperature range 0 to 50 °C

Calibrated according to JIS C-1602-2015 standard thermoelectric power mV input

Resistance Temperature Detector (RTD)
Code Temperature sensor Measuring Range Display Range Accuracy *2
R Pt 100Ω -199.9 to 199.9°C -199.9 to 199.9°C ±(0.2% of rdg. + 0.3°C)
G -199 to 850°C -199 to 870°C ±(0.4% of rdg. + 1°C)
*2 Accuracyy:
Specified at measurement range
Defined at 23 ± 5°C, 45 to 75% RH.
Temperature coefficient:
±200 ppm/°C for 0 to 50 °C

Calibrated according to JIS C-1604-2013 standard resistance element resistance values

②Display Color

Code Content
R Red LED
Green LED

③Option

Code Content
X none
A01 Brightness can be adjusted.

General Specifications

Display Red or green LED (character height 10 mm) with zero-suppression function.
Over-range display When exceeding display range, displays minimum or maximum value flashing.
Open circuit of input Thermocouple input: Displays minimum value of display range flashing.
RTD input: Displays maximum value of display range flashing
Resolution Thermocouple input: 1°C
RTD input: 1°C or 0.1°C
External resistance Thermocouple input: Less than 500 Ω
RTD input: Lead wire 1 line less than 5 Ω
Powerload DC ±3.3 V
Sampling cycle 2.5 times/sec.
Input configuration Single ended
A/D conversion Δ-Σ conversion system
Noise reduction Normal mode (NMR)  50dB or higher
Hold function Holds measured data and data output. (Not isolated from the input)
Withstand voltage Between input terminal and external case: AC 500 V 1minute
Between power supply terminal and external case: AC 500 V 1minute
Between power supply terminal and input terminal: AC 500 V 1minute
Insulation resistance DC 500 V 100 MΩ or more
Power supply DC 12 to 24 V
Power supply voltage tolerance DC 10.8 to 32 V
Power consumption DC 12 V … 32 mA
DC 24 V … 20 mA
Operating ambient temperature 0 to 50°C
Storage temperature -20 to 70°C
Weight Approx. 40 g

page top