Our Technology, solution for Tomorrow too!

Measuring Instrument & Tester

AC mΩ Meter

MODEL 358E

  • ■Suitable for measurement of resistance and voltage of Fuell cell
  • ■Measuring range 30 mΩ, 300 mΩ and 3 Ω with measuring frequency 10 kHz
  • ■Variation of output, RS-232C and analog outputs for resistane and voltage
  • ■Optional BCD output is also available

Catalog
Operation manual

Model Designation

358E -
 
① Data Output
Code Specification
X RS-232C + analog output (standard)
03 BCD (TTL output) + RS-232C + analog output
04N BCD (NPN output) + RS-232C + analog output
04P BCD (PNP output) + RS-232C + analog output

Measuring range

【Resistance Measurement】

Sampling speed: SLOW and MEDIUM
Measurement Range 30 mΩ 300 mΩ 3 Ω
Resolution 1 μΩ 10 μΩ 100 μΩ
Measurement Current 7.4mA 1 mA 100 μA
Accuracy* ± (1.0 % of rdg.+ 20 digit) ± (1.0 % of rdg.+ 15 digit)
Temperature coefficient ± (0.1 % of rdg.+ 2 digit)/°C ± (0.1 % of rdg.+1digit)/°C
Open terminal voltage 1 V peak or less
*Accuracy:
Specified at 23 ℃ ± 5 ℃ 45 to 75 % RH
Sampling speed: SLOW and MEDIUM mode
Measurement Range 30 mΩ 300 mΩ 3 Ω
Resolution 1 μΩ 10 μΩ 100 μΩ
Measurement Current 7.4mA 1 mA 100 μA
Accuracy* ± (1.0 % of rdg.+ 15 digit) ± (1.0 % of rdg.+ 5 digit)
Temperature coefficient ± (0.1 % of rdg.+ 1.5 digit)/°C ± (0.1 % of rdg.+1digit)/°C
Open terminal voltage 1 V peak or less
*Accuracy:
Specified at 23 ℃ ± 5 ℃ 45 to 75 % RH
【Voltage measurement】
Sampling cycle:  SLOW, MEDIUM and FAST mode
Measuring range DC ± 5 V DC ± 50 V
Resolution 100 µV (1 mV) 1 mV (10 mV)
Accuracy ± (0.05% of rdg. + 5 digit.)*
Temperature coefficient ± (0.005% of rdg. + 0.5 digit.)/℃
Accuracy:
Specified at 23 ℃ ± 5 ℃, 45 to 75 %RH
3 digits added for sampling MEDIUM
5 digits added for sampling speed FAST

General Specifications

Measuring method AC 4-terminal method
Maximum allowable applied voltage All ranges 60 V DC
Measurement frequency AC 10 kHz ± 20 Hz
Lead Wire Resistance SOURCE Hi and SOURCE Lo 0.4Ω or less
Display OLED display
Resistance measurement: 35000 or 3500
Voltage measurement: 50050 or 5050 (with negative sign)
With Zero suppression function
Over range display OVER, UNDER
Sampling cycle SLOW: 2.5 times/s
MEDIUM: 5 times/s
FAST: 60 times/sec, 50 times/s
Response speed SLOW: approx. 1.2s
MEDIUM: approx. 600 ms
FAST: approx. 50 ms (at sampling cycle 60 times/s)
Insulation resistance Between power supply terminal and Outer case: DC 500 V More than 100 MΩ
Between mesuring terminal and Control terminal: DC 500 V More than 100 MΩ
Withstand voltage Measuring terminal, Control terminal, RS-232C and Outer case: AC 500 V 1 minute
Between power supply terminal and Outer case: AC 1500 V 1 minute
Parameter retention Parameters are saved and retained in non-volatile memory.
Power supply AC 100 to 240 V (50/60 Hz)
Power consumption Approx. 10 VA, Approx. 17 VA
Ambient operation temperature 0 to 40 °C, 80 % RH or less (no condensation)
Storage temperature -20 to 65 °C, 80 % RH or less (no condensation)
Weight Approx. 1.3 kg
Size 205(W) x 65(H) x 169(D) mm
Accessory Power cord 1 pc
Operating instructions 1 pc
Utility software (CD) 1pc
Optional Accessories Kelvin clip 5811-24B
Foot switch for utility software 5858-04U
RS-232C to USB conversion cable 5858-20
Panel-mount bracket 5811-31

page top