Our Technology, solution for Tomorrow too!

Digital Meter

Digital Large Waterproof Display Meter

MODEL 4027, 4028

  • ■ Smart design
  • ■ Safe waterproof type. (IP65)
  • ■ 60mm high brightness dot matrix LED for clear display
  • ■ Wall or hanging-mount available.

Catalog

Operation manual

Standard Specifications

Model Designation

402 □-□ - - -
  └①┘      

①Measuring Input

Code Measuring Range
(type of input)
Display Range/Input Resistance Accuracy*
7-02 -200 to 999°C(K) -270 to 999°C ±(0.3% of rdg. + 1°C)
7-04 40 to 750°C(J) -210 to 999°C ±(0.3% of rdg. + 1°C)
7-05 -200 to 350°C(T) -270 to 420°C ±(0.3% of rdg. + 1°C)
7-11 -200 to 850°C(PT100Ω) -200 to 870°C ±(0.4% of rdg. + 1°C)
7-12 -99.9 to 99.9°C(PT100Ω) -99.9 to 99.9°C ±(0.2% of rdg. + 0.3°C)
8-03 ±999 mV(DC V) 1 MΩ ±(0.1% of rdg. + 1 digit)
8-04 ±9.99 V(DC V) 1 MΩ ±(0.1% of rdg. + 1 digit)
8-09 1 to 5 V 1 MΩ ±(0.1% of rdg. + 2 digit)
8-19 4 to 20 mA 12.5 Ω ±(0.1% of rdg. + 2 digit)
*Accuracy:
Defined at 23 ± 5°C, 45 to 75% RH.
Temperature coefficient:
± 300 ppm/°C (K,J,T), ± 200 ppm/°C (Pt), ± 100 ppm/°C (-03), ± 150 ppm/°C (-04, -09, -19)
Operating temperature range:
-10 to 50 °C
Overload:
DC ± 10 V (K,J,T,Pt), DC ± 250 V (-03, -04, -09), DC ± 150 mA (-19)
Calibration:
JIS C-1602-2015 standard thermoelectric power mV (K,J,T)

JIS C-1604-2013 standard resistance element resistance values (Pt)

②Power Supply

Code Content
A AC 100 to 240 V
9 DC 24 V

③Mounting method / Sensor (only for RTD input)

Code Mounting method Sensor Protection rate
51 Wall-hanging Sensor not included IP55 equivalent
52 Suspending
53 Wall surface mounting IP65 equivalent
61* Wall-hanging Integrated sensor type IP55 equivalent
62* Suspending
63* Wall surface mounting IP65 equivalent

* Set with sensor

④Optional Function

Code Function
X Peak hold
A01 Bottom hold
A02 Burnout (min. value flashes)*

* Sensors of K,J,T only

General Specifications

Display 3 digits  0 to 999
High brightness large LED (character height 60 mm) in Red
Zero suppression function is equipped.
Polarity display (POL) " - " is displayed.
Over-range display Temperature input: Min. or Max. value of display range flashes.
DC input: Value of 130% of display range flashes. But 000 flahes, when it exceeds 999.
Decimal point Temperature input: Fixed point
DC input: Arbitrary setting
Burnout Max. value of display range flashes. (only for temperature input)
Resolution Thermocouple input: 1 ℃
RTD input: 1 ℃ or 0.1 ℃
DC input: 1/1000
External resistance Thermocouple input: less than 500 Ω
RTD input: Lead wire line less than 5 Ω
Display cycle Temperature input: Approx. 5 times/s.
DC input: Approx. 15 times/s.
Protection rate Wall-hanging and suspending installation: IP55 equivalent
Wall surface mounting installation: IP65 equivalent
Input configuration Single ended input
A/D conversion Δ-∑ conversion system
Noise rejection rate Normal mode (NMR): more than 50 dB
Common mode (CMR): more than 110 dB
Power line mixed noise: 1000 V
Insulation resistance DC 500 V  More than 50 MΩ
Withstand voltage Between input terminal and outer case: AC 1500 V 1min.
Between power supply terminal and outer case: AC 1500 V 1min.
Between power supply terminal and input terminal: AC 1500 V 1min.
(DC power supply model: AC500 1min.)
Power supply AC 100 to 240 V, (50/60Hz)
DC 24 V
Power supply voltage tolerance AC 85 to 250 V
DC 20 to 30 V
Power consumption AC 100 V approx. 7 VA
AC 200 V approx. 9 VA
DC 24 V approx. 100 mA
Operating ambient temperature -10 to 50°C
Storage temperature -20 to 70°C
Weight Approx. 1.5 kg
Unit display Attached on front panel. ("℃" unit is attachd for 4027, Required unit can be attched for 4028, please specify at the order.)