Transducer
Basic Specifications
Model Designation
| 7522 | - | □ | - | □ | - | □ |
| ① | ② | ③ |
| Code | Input |
|---|---|
| 1E | R 0 to 1400°C * Accuracy guarantee range :400 to 1400°C |
| 28 | K 0 to 300°C |
| 29 | K 0 to 400°C |
| 2A | K 0 to 500°C |
| 2B | K 0 to 600°C |
| 2C | K 0 to 1000°C |
| 2D | K 0 to 1200°C |
| 37 | E 0 to 200°C |
| 45 | J 0 to 100°C |
| 46 | J 0 to 150°C |
| 47 | J 0 to 200°C |
| 49 | J 0 to 400°C |
| 55 | T 0 to 100°C |
| 56 | T 0 to 150°C |
| 57 | T 0 to 200°C |
| 58 | T 0 to 300°C |
| Code | Output |
|---|---|
| 01 | DC 0 to 10 mV |
| 02 | DC 0 to 100 mV |
| 03 | DC 0 to 1 V |
| 04 | DC 0 to 5 V |
| 05 | DC 0 to 10 V |
| 09 | DC 1 to 5 V |
| 00 | Except above DC voltage output (More than 10 mV less than 10 V) |
| 13 | DC ± 1 V |
| 14 | DC ± 5 V |
| 10 | Except above DC ± voltage output (More than ± 10 mV less than ± 10 V) |
| 23 | DC 0 to 1 mA |
| 29 | DC 4 to 20 mA |
| 20 | Except above DC current output (More than 100 µA less than 20mA) |
| Code | Power |
|---|---|
| 3 | AC 100/110 V |
| 5 | AC 200/220 V |
| 9 | DC 24 V |
Output Specifications
| Output Signal | Output Impedance | Tolerable Load Resistance |
|---|---|---|
| DC 0 to 10 mV | Approx. 10 Ω | More than 10 kΩ |
| DC 0 to 100 mV | Approx. 100 Ω | More than 100 kΩ |
| DC 0 to 1 V | Less than 0.1 Ω | More than 100 Ω |
| DC 0 to 5 V | More than 500 Ω | |
| DC 0 to 10 V | More than 1 kΩ | |
| DC 1 to 5 V | More than 500 Ω | |
| DC ± 1 V | ||
| DC ± 5 V | More than 2.5 kΩ | |
| DC 0 to 1 mA | More than 5 MΩ | 0 to 12 kΩ (During DC Power 0 to 10 kΩ) |
| DC 4 to 20 mA | 0 to 600Ω (During DC Power 0 to 500 Ω) |
General Specifications
| Input | Selectable thermocouple Tolerable external resistance: Less than 500 Ω |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Allowance | ±0.3% of SPAN at 23°C | |||||||||
| Cold junction compensation accuracy | sensor K, E, J or T: Less than ± 0.5°C sensor R: Less than ± 1°C (under the condition of 0 to 40°C) |
|||||||||
| Response | Less than 0.5s (0→90%) | |||||||||
| Temperature coefficient | 200 ppm/°C | |||||||||
| Output adjustment range | ZERO:More than ± 3% of SPAN SPAN:More than ± 5% of SPAN |
|||||||||
| Insulation resistance | Between Input terminals, Output terminals and Power supply terminals : DC 500 V More than 100 MΩ | |||||||||
| Withstand voltage | Between Input terminals and Power supply terminals: AC 1500 V 1min.
(During DC Power: AC 1000 V 1min.) Between Input terminals and Output terminals: AC 1500 V 1min. (During DC Power: AC 1000 V 1min.) Between all terminals and external case: AC 1500 V 1min. |
|||||||||
| Power supply voltage |
|
|||||||||
| Operating temperature | -5 to 55°C | |||||||||
| Storage temperature | -20 to 70°C | |||||||||
| Weight |
|