Measuring Instrument & Tester

Withstand Voltage and Insulation Tester

MODEL 8525A

  • ■Withstand voltage test: AC5kV 500VA, Insulation resistance testing: DC500V / 1000V
  • ■Safety measurements like Interlocks, Key locks and Remote function
  • ■RS-232C output enables easy setting and obtaining test results.
  • ■ 9 types of test memories can be stored.
  • ■JCSS calibration available at request

Catalog
Operation manual

Model Designation

8525A                    : for power supply voltage AC100V (Standard)
8525A-P115           : for power supply voltage AC115V
8525A-P200           : for power supply voltage AC200V
8525A-P220           : for power supply voltage AC220V
8525A-P240           : for power supply voltage AC240V

Specifications

Withstand voltage Test

Test voltage
Output Voltage AC 0 to 2.5 kV / 0 to 5 kV, 2 ranges
Output Capacity 500 VA (5 kV, 100 mA)
Note: For output current 50 mA within 30 min.
Waveform Commercial power supply waveform
Voltage Fluctuation Rate 15 % or less (no load → maximum load)
Voltage Output Method Zero cross input switch
Output Voltage Setting Manual setting with volt slider
Voltage Measurement
Rectification Method Average value effective value display
Analog Indicator Full scale: AC 0 to 5 kV
Accuracy ± 5 % (JIS C1102)
Digital Indicator Display range: AC 0.00 to 6.00 kV Green LED
Output voltage is displayed during test, the judgment value is retained after the end of test.
Measurement Accuracy ± 1.5 % of F.S. (at F.S. 2.5 kV or 5 kV)
Current Measurement
Rectification Method Average value effective value display
Measuring Range ± 1.5 % of F.S. (at F.S. is 2.5 kV or 5 kV)
Display Digital 3.5 digit display (1999) Green LED
Leakage current is displayed during test, the judgment value is retained after the end of test.
Resolution 0.01 mA (0.1 to 9.9 mA), 0.1mA (10.0 to 110.0 mA)
( ) indicates upper limit setting value
Measuring Accuracy Upper limit setting value ± (5 %+20 μA)
Test Result Judgment
Judgment Method Upper limit Analog comparator
Upper/lower limit Digital comparator
Setting Range Upper limit: 0.1 to 110.0 mA (lower limit setting + More than 1 digit)
Lower limit: 0.0 to 109.0 mA (upper limit setting - Less than 1 digit)
Resolution: 0.1 mA with lower limit ON/OFF setting function
Judgment Conditions Upper limit > leakage current > lower limit ... GOOD
Upper limit ≦ leakage current ... HIGH NG
Lower limit ≧ leakage current ... LOW NG
Test Time
Setting Range 0.5 to 999 s (with Timer OFF function)
Setting Resolution 0.1 s (0.5 to 99.9 s), 1 s (More than 100 s)
Time Display Timer ON: Remaining time is display
Timer OFF: Elapsed time is display
Accuracy ± 20 ms (0.5 to 99.9 s), ± 200 ms (More than 100 s)

Insulation Resistance Test

Test Voltage
Rated Measurement Voltage DC 500 V / DC 1000 V 2 range switching
(DC 25 V / 50 V / 100 V / 250 can be manufactured per request.)
Open Circuit Voltage Within 125% of rated measurement voltage
Rated Measurement Current 1 mA
Short-circuit Current Less than 12 mA
Display 0.00 to 2000 MΩ Green LED
Insulation resistance is displayed during test, the judgment value is retained after the end of test.
Measurement Range and Accuracy
[Rated Measurement Voltage] [Display Range] [Resolution] [Accuracy]
DC 500V,
DC 1000 V
0.00 to 20.00 MΩ 10kΩ ±(2 % of rdg. +3digit)
18.0 to 200.0 MΩ 100 kΩ
180 to 2000 MΩ 1MΩ ±(5% of rdg. +3digit)
Accuracy: specified at 23 °C ± 5 °C, 45 to 75 %RH
Test Result Judgment
Judgment Method Upper/lower limit digital comparator
Setting Range
Upper limit:
0.2 to 2000 MΩ (lower limit setting +More than 1digit)
Lower limit:
0.1 to 1999 MΩ (upper limit setting - Less than 1digit)
Resolution:
0.1 MΩ (0.2 to 9.9 MΩ, lower limit range is 0.1 to 9.11 MΩ)
1MΩ (10 to 2000 MΩ, lower limit range is 10 to 2001 MΩ)
With upper limit ON/OFF function
Judgment Conditions Upper limit > display value > Lower limit ... GOOD
Upper limit ≦ display value... HIGH NG
Lower limit ≧ display value ... LOW NG
Test Time
Setting Range 0.5 to 999 s
(Mask time +0.2 s or more, with Timer OFF function)
Setting Resolution 0.1 s (0.5 to 99.9 s), 1 s (100 to 999 s)
Time Display During test
Timer ON: Remaining time display
Timer OFF: Elapsed time display
Mask Time
Setting Time 0.3 to 50.0 s (test time -0.2s or less)
Accuracy ± 20 ms (0.5 to 99.9 s), ± 200 ms (100 to 999 s)
Time Display During test
Timer ON: Remaining time display
Timer OFF: Elapsed time display
Discharge Function
Electric charge on specimen can be discharged.
(with OFF function)

Data Output

Voltage Monitor Output ... Output voltage can be obatained for monitoring.
Monitor Output DC 0 to 5 V (corresponding to AC 0 to 5 kV)
Accuracy ± 1.5 % of F.S.
RS-232C Output ... Test conditions and test results can be output. ※Utility software (option) is available.
Connector D-sub 9P
Transmission Method Full duplex synchronization
Transmission Speed 9600 bps
Data Length 8 bit
Parity None

General Specifications

Power Supply AC100V 50/60Hz (DC 25 V / 50 V / 100 V / 250 are available.)
Power Supply Voltage Tolerance AC 90 to 110 V 50 / 60 Hz
Power Consumption Maximum rated load: approx. 650 VA
No load (standby state): approx. 16 VA
Operating Ambient Temperature 0 to 40 °C
Operating Humidity Range Less than 80 %RH (no condensation)
Storage Temperature -20 to 65 °C Less than 80 %RH (no condensation)
Insulation Resistance Between power supply and outer box AC1000V 1 minute
Withstand Voltage Between power supply and outer box AC1000V 1 minute
Size 320(W) × 165(H) × 365(D )mm
Weight Approx. 17 kg
(Approx. 5.5kg is added for optional power supply models.)
Accessories High voltage cable: 2m 1 set
Ground cable: 3m 1 pc
Power cord: 2.5m 1 pc
REMOTE I/O plug: 1 pc
Fuse: 1 pc
Operation manual: 1 pc
Optional Accessories Remote control box 5858-07
Both-hands remote control box 5858-07W
Foot switch 5858-04
Rack mounting bracket 5871-03-020
RS-232C cable 5881-11-018
Utility software (Japanese) 5890-02