Our Technology, solution for Tomorrow too!

Measuring Instrument & Tester

10 kV type AC Withstand Voltage Tester

MODEL 8529A

  • ■Output voltage: AC 10 kV, 500 VA
  • ■9 types of test memories can be stored.
  • ■RS-232C output enables easy setting and obtaining test results.
  • ■Safety measurements like Interlocks, Key locks and Remote function are available.

Catalog
Operation manual

Model Designation

8529A                    : for power supply voltage AC100V (Standard)
8529A-P115           : for power supply voltage AC115V
8529A-P200           : for power supply voltage AC200V
8529A-P220           : for power supply voltage AC220V
8529A-P240           : for power supply voltage AC240V

Specifications

Test Voltage
Output Voltage AC 0 to 5 kV / 0 to 10 kV, 2 ranges
Output Capacity 500 VA (10 kV, 50 mA)
Note: For output current 25 mA within 30 min.
Waveform Commercial power supply waveform
Voltage Fluctuation Rate Less than 15 % (no load → maximum load)
Voltage Output Method Zero cross input switch
Output Voltage Setting Manual setting with volt slider
Voltage Measurement
Rectification Method Average effective value display
Analog Indicator Scale: AC 0 to 10 kV
Accuracy: ± 5 %
Digital Indicator Display range: AC 0.00 to 12.0 kV Green LED
Output voltage is displayed during test, the judgment value is retained after the end of test.
Accuracy: ± 1.5 % of F.S. (at F.S. 5 kV or 10 kV)
Current Measurement
Rectification Method Average effective value display
Measuring Range 0.01 to 55.0 mA
Display 3.5 digit (1999) digital display Green LED
Leakage currentis displayed during test, the judgment value is retained after the end of test.
Resolution AC 0.01 mA (0.1 to 9.9 mA), 0.1mA (10.0 to 55.0 mA)
( ) indicates upper limit setting value
Measuring Accuracy Upper limit setting value ± (5 %+20 μA)
Test Result Judgment
Judgment Method Upper limit: Analog comparator
Upper/lower limit: Digital comparator
Setting Range Upper limit: 0.1 to 55.0 mA
Lower limit: 0.0 to 54.9 mA
With ON/OFF function for lower limit
Resolution: 0.1mA
Judgment Conditions Upper limit > display value > Lower limit ... GOOD
Upper limit ≦ display value ... HIGH NG
Lower limit ≧ display value ... LOW NG
Test Time
Setting Range 0.5 to 999 s (with Timer OFF function)
Setting Resolution 0.1 s (0.5 to 99.9 s), 1 s (100 to more than 999 s)
Time Display Timer ON: Remaining time is display
Timer OFF: Elapsed time is display
Accuracy ± 20 ms (0.5 to 99.9 s), ± 200 ms (100 to 999 s or more)

Data Output

Voltage Monitor Output ... Output voltage can be obatained for monitoring.
Monitor Output DC 0 to 5 V (corresponding to AC 0 to 5 kV)
Accuracy ± 1.5 % of F.S.
RS-232C Output ... Test conditions and test results can be output.
※Utility software (option) is available.
Connector D-sub 9P
Transmission Method Full duplex synchronization
Transmission Speed 9600 bps
Data Length 8 bit
Parity none

General Specifications

Power Supply AC100V 50/60Hz
(AC 115 V / 200 V / 220 V / 240 are available.)
Power Supply Voltage Tolerance AC 90 to 110 V
Power Consumption Rated load: approx. 650 VA
No load (standby state): approx. 16 VA
Operating Ambient Temperature 0 to 40 °C
Operating Humidity Range 20 to less than 80 %RH (no condensation)
Storage Temperature -20 to 70 °C Less than 90 %RH (no condensation)
Withstand Voltage Between power supply and Outer box: AC 1000 V 1 minute
Size 320(W) × 180(H) × 430(D )mm
Weight Approx. 20 kg
(Approx. 5.5kg is added for optional power supply models.)
Accessories High voltage cable: 2m 1 set
Ground cable: 3m 1 pc
Power cord: 2.5m 1 pc
REMOTE I/O plug: 1 pc
Fuse: 1 pc
Operation manual: 1 pc
Optional Accessories Remote control box 5858-07
Both-hands remote control box 5858-07W
Foot switch 5858-04
Rack mounting bracket 5871-03-020
RS-232C cable 5881-11-018
Utility software (Japanese) 5890-02

page top